Calibration

Calibrate a channel so its meters read dB SPL: measure a calibrator on the capsule, compare against a trusted meter, or enter a known offset.

A calibration is an offset between what the interface reports in dBFS and the real sound pressure level at the mic. Once a channel has one, its meter, the DB Only tile and the cursor readout show dB SPL as well as dBFS.

  1. Start a capture on the device and channel count you’ll measure with. See Getting audio in.
  2. Click Calibrate… on the channel’s meter.

The Calibrate Ch <n> dialog opens. Current status shows valid if the channel has a calibration that still applies, or why not, for example that the channel has never been calibrated.

The dialog has three tabs.

The preferred method, and the one you can trace.

  1. Put the calibrator on the mic capsule and switch it on.
  2. Choose its level: 94 dB SPL or 114 dB SPL.
  3. Click Start 5s measurement.

While it measures, the dialog shows the Live level and the Stability (spread): keep the calibrator steady so the spread stays small. Then it shows what it Measured and the resulting offset.

If the meter didn’t deliver enough readings in the five seconds, the dialog says the measurement can’t be trusted. Hold the calibrator steady and measure again.

Click Save calibration.

Calibrate against a sound level meter you trust, on the same source.

  1. Set your meter to the same weighting and time constant as the RTA. The dialog tells you what the RTA is using, for example Z-weighted, slow. A mismatch gives an offset that’s wrong by the difference between the weightings.
  2. Enter the meter’s reading in Trusted meter reading: (dB SPL). The dialog shows App’s current reading next to it.
  3. Click Save calibration.

Restore an offset you already know, from an earlier session or an outside measurement.

  1. Enter Offset (dB):.
  2. Enter the Reference level used (dB SPL):.
  3. Click Save calibration.

A manual entry is recorded as a comparison. It’s still tied to the capture that’s running now, so it goes stale like any other calibration.

A calibration belongs to the device, sample rate and channel count it was made with. If any of those changes, for example you start on a different interface or with a different number of channels, the channel’s SPL is hidden and the meter says why. dBFS keeps working. Go back to the original setup, or calibrate again.

SPL is also hidden while the browser reports its voice processing on. See Getting audio in.

To remove a calibration, open the dialog and click Clear calibration.